62
3.2 Basic UV-VIS-IR Absorption, Emission, and Elastic Light Scattering Methods
Conventional FTIR is limited by a combination of factors including an inevitable trade-off
between acquisition time and SNR and the fact that there is no spatial localization informa
tion. However, recent developments have utilized multiple intense, collimated IR beams
from synchrotron radiation (see Chapter 5). This approach permits spatially extended detec
tion of IR absorption signals across a sample allowing diffraction-limited time-resolved high-
resolution chemical imaging, which has been applied to tissue and cellular samples (Nasse
et al., 2011).
Many biomolecules that contain chemical bonds that absorb in the IR will also have a
strong Raman signal. The Raman effect is one of the inelastic scattering of an excitation
photon by a molecule, resulting in either a small increase or decrease in the wavelength of the
scattered light. There is a rough mutual exclusion principle, in that strong absorption bands
in the IR correspond to relatively weak bands in a Raman spectrum, and vice versa. Raman
spectroscopy is a powerful biophysical tool for generating molecular signatures, discussed
fully in Chapter 4.
Long UV light (~200–400 nm) is also a useful spectrophotometric probe especially for
determining proteins and nucleic acid content in a sample. Peptide bonds absorb most
strongly at ~280 nm wavelength, whereas nucleic acids such as RNA and DNA have a peak
absorption wavelength of more like ~260 nm. It is common therefore to use the rate of absorp
tion at these two wavelengths as a metric for protein and/or nucleic acid concentration. For
example, a ratio of 260/280 absorbance of ~1.8 is often deemed as “pure” by biochemists for
DNA, whereas a ratio of ~2.0 is deemed “pure” for RNA. If this ratio is significantly lower, it
often indicates the presence of protein (or potentially other contaminants such as phenol that
absorb strongly at or near 280 nm). With suitable calibration, however, the 260 and 280 nm
absorption values can be used to determine the concentrations of nucleic acids and proteins
in the absence of sample contaminants.
In basic spectrophotometers, the transmitted light intensity from the sample is ampli
fied and measured by a photodetector, typically a photodiode. More expensive machines will
include a second reference beam using an identical reference cuvette with the same solvent
(generally water, with some chemicals to stabilize the pH) but no sample, which can be used as
a baseline against which to reference the sample readings. This method finds utility in meas
uring sample density containing relatively large biological particulates (e.g., cells in suspension,
to determine the so-called growth stage) to much smaller ones, such as molecules in solution.
To characterize attenuation, if we assume that the rate absorption of light parallel to the
direction of propagation, say z, in an incrementally small slice through the sample is propor
tional to the total amount of material in that thin slice multiplied by the incident light inten
sity I(z), then it is trivial to show for a homogeneous tissue:
(3.3)
I z
I
Cz
( ) = ( )
−
(
)
0 exp
(
σ λ
)
This is called the “Beer–Lambert law,” a very simple model that follows from the assumption
that the drop in light intensity upon propagating through a narrow slice of sample is
TABLE 3.1 Common Covalent Bonds in Biological Matter, Which
Absorb Infrared Electromagnetic Radiation, with
Associated Typical Absorption Ranges Indicated
Peak IR Absorption Range (cm−1)
Bond in Biomolecule
730–770
C—H
1180–1200
C—O—C
1250–1340
C—N
1500–1600
C=C
1700–1750
C=O
2500–2700 (and other peaks)
O—H
3300–3500 (and other peaks)
N—H